Chapter 3: Getting Started
Product Service 1-800-522-7658
38
Original instructions
Audible alarm Mgmt.
Arrow down to “Fan set point” press “Ok” to select it.
You should now see:
>Default (2850)
Minimal (1500)
Reference your GFH Settings Sheet, arrow down to “Default” press “OK” to
select it. Enter value from GFH Settings Sheet.
“Confirm new value” press “Ok” to confirm.
Arrow down to “Minimal” press “Ok” to select it.
Enter value from GFH Settings Sheet.
“Confirm new value” press “Ok” to confirm.
Press “X” to return to the previous menu.
Step 11 Set alarm delays and choose the sash type.
Reference your GFH Settings sheet for the following step:
You should see:
>Hood integration
Fan set point
Semicon S. Alarm Delay
Electro S. Alarm Delay
Sash sensor trigger
Software releases
Audible alarm Mgmt.
Arrow down to “Semicon S. alarm Delay” press “Ok” to select it.
You will see:
05 Minutes
1 Min
60 Max
Make the 0 a 1 for 15 Minutes
Press “Ok” then “Confirm new value” press “Ok” to confirm.
You should see this menu again:
>Hood integration
Fan set point
Semicon S. Alarm Delay
Electro S. Alarm Delay
Sash sensor trigger
Software releases
Audible alarm Mgmt.
Arrow down to “Electro S. alarm Delay” press “Ok” to select it.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Protector Echo 180 Series and is the answer not in the manual?

Labconco Protector Echo 180 Series Specifications

General IconGeneral
BrandLabconco
ModelProtector Echo 180 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals