Chapter 3: Getting Started
Product Service 1-800-522-7658
39
Original instructions
You will see:
05 Minutes
1 Min
60 Max
Make the 0 a 1 for 15 Minutes
Press “Ok” then “Confirm new value” press “Ok” to confirm.
You should see this menu again:
>Hood integration
Fan set point
Semicon S. Alarm Delay
Electro S. Alarm Delay
Sash sensor trigger
Software releases
Audible alarm Mgmt.
Arrow down to “Sash sensor trig.” Press “Ok” to select it.
Refer back to your notes from step 2 for the type of Sash sensor and trig. X%
value
Arrow down to “SRS Inverted Polarity” then press “Ok” to select it.
“Confirm new value” press “Ok.”
You will see Sash sensor trigger.
Enter the 65% as the trig value, press “Ok.”
“Confirm new value” press “Ok.”
Press “X” to return to the previous menu.
Step 12 Authorize Administrator for alarms.
You should see:
>Hood integration
Fan set point
Semicon S. Alarm Delay
Electro S. Alarm Delay
Sash sensor trigger
Software releases
Audible alarm Mgmt.
Arrow down to “Audible alarm Mgmt.” Press “Ok” to select it.
You should see “Admin Authorized No. Press “Ok” to have it say “Yes.
Arrow down to save and exit press “Ok.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Protector Echo 180 Series and is the answer not in the manual?

Labconco Protector Echo 180 Series Specifications

General IconGeneral
BrandLabconco
ModelProtector Echo 180 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals