Chapter 3: Getting Started
Product Service 1-800-522-7658
40
Original instructions
Step 13 Set audible alarms.
You should now see:
Hood Status
Administration menu
Maintenance menu
Arrow up to “Administration menu” and press “Ok” to select it.
You should now see:
Language
Date and Time
Alarms and Sensors
Users and Management
Network Parameters
Erase History
Ventilation
Arrow down to “Alarms and Sensors. Press “Ok” to select it.
You should see:
>Maintenance Alarm
Deactivation Rights
Semicon S. Trig. Value
Sash Sensor Calib
Audible alarm activation
Arrow down to “Audible alarm activation. Press “Ok” to select it.
Turn on alarms that are stated on the GFH settings sheet. For example 8K, 8N,
8Q, 8V, and 8W.
Arrow down to 8K. Press “OK” to select it.
You should see “Repl. Main filters: Off. Press “Ok” to make it a “Yes.
Then arrow down to “save and exit. Press “Ok” to save and exit.
Repeat for all the remaining alarms.
After the last alarm is set, press “X” to return to the previous menu.
Step 14 Calibrate the sash.
You should see:
Maintenance Alarm
Deactivation Rights
Semicon S. Trig. Value
Sash Sensor Calib
Audible alarm activation

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Protector Echo 180 Series and is the answer not in the manual?

Labconco Protector Echo 180 Series Specifications

General IconGeneral
BrandLabconco
ModelProtector Echo 180 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals