Chapter 3: Getting Started
Product Service 1-800-522-7658
42
Original instructions
Example, M4 will have four fans. Your numbers will be different based on your
parameters.
Fan 1; 1533
Fan 2: 1522
Fan 3: 1511
Fan 4: 1630
Fan 5: -------
Example
Semicon Sensor
Trig. M 3500 MV
Delay 15mn
Electro Sensor
Delay 15 mn
Sash
Type Inverted polarity
Trig. 65%
Now check the audible alarm for the sash by raising the sash so it sounds. The
height you hear it should be the upper end of the safe working height above the
sash stop.
If not, make sure you calibrated it. If needed, recalibrate it until it is correct. To do
this turn the hood off and reselect the sash type and then recalibrate it.
They should all match your notes from step 2 and your GFH settings sheet except
that the days will start over at 0 and count to 730 or the number that was entered.
***If everything was entered correctly, the filtered hood will function properly.
Otherwise you must turn it off and reenter the correct parameters. If you still have
a problem please contact Labconco Product Service.***

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Protector Echo 180 Series and is the answer not in the manual?

Labconco Protector Echo 180 Series Specifications

General IconGeneral
BrandLabconco
ModelProtector Echo 180 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals