CoaDATA 501 – Operators Manual 2.1 Page 8
1.5 Reagents
For proper coagulation analysis we recommend to use reagents, controls
and buffers from well known reagent manufacturers.
Contamination
With the application of different reagents, and here especially reagents
containing thrombin, there is a danger of reagent carry-over.
When adding reagents the light protection cap is exposed to
reagents and hence a point of contamination.
This point of contamination must be cleaned with a suitable thrombin in
activator and a cotton swab after each method change.
Useful hints:
Use the pipettor supplied with the respective pipette tips (Thrombi-Tips).
In General: The analyzer is equipped with light protection caps for
Thrombi-Tips.
To guarantee perfect functioning of the system it is absolutely
necessary to use only the variable Thrombi-Pette and original
Thrombi-Tips. Please understand that warranty shall not apply to
instruments which have problems due to the use of other types of
pipettes or tips.
Make sure no air bubbles are generated during the pipetting process.
Read the reagent packaging insert prior to use and follow the instructions.
Only use original cuvettes and stir bars from the manufacturer which
are subject to strict quality control measures. Please understand
that the use of non-original cuvettes or instruments problems
caused by the use of non-original cuvettes may lead to the warranty
obligation become null and void.
Perform analytical quality controls on a regular basis.