Operation
CoaDATA 501 – Operators Manual 2.1 Page 25
2.2.3 Fibrinogen 1 [g/l] - parameterization
• Select method < 3 Fib. 1 >
< 3 Fib. 1 >
Results will be calculated in [g/l]
• Press Mode; and enter the secret no.
<- secret no.:
: ______
(preset to: 11111)
If you enter the wrong number STANDBY will appear.
The following dialog will appear as soon as the correct number has been
entered:
< 1.conv >
The menus <1. conversion>, <replication>, <measurement> and <cuv
remove detec.> will be displayed by pressing the right arrow key.
Overview over Fibr. g/l parameterization
<1.conversion>
<curve>
input of a 9-point calibration curve under or
<none>
no conversion
<replication>
select single or double determination
<single>
and the coefficient of variation
<double>
(CV 1-20%).
<measurement>
start reagent volume, reagent lot. no.
and sample incubation time.
<cuv remove detect.>
refer to chapter 3 Software
<ON>
activates the automatic cuvette detection
<OFF>
deactivates the automatic cuvette detection.
(default)
• Press Enter to confirm the selection.
1st conversion
A 9-point calibration curve can be entered with this menu.
Calibration curve points that are not to be used must have the entry 0.0 s.
For information on interpolation refer to chapter 5.5 Mathematics. The
system requires two points to be defined but we would recommend a
minimum of 3 points. You can exit the calibration curve menu with Esc if no
entry has been made. Once an entry has been made, all additional
calibration curve points must be retrieved and verified.
< curve >
• Press Enter to enter the calibration curve.
1.point:
Definition as: