Operation
CoaDATA 501 – Operators Manual 2.1 Page 32
As soon as the temperature has stabilized at 37.4 °C on the display intern
the value must be compared with the value measured with the digital
multimeter. If the internal and external values are identical temperature
adjustment has been completed. Otherwise the procedure has to be
repeated.
• Press Esc. All values will be stored by the analyzer.
2.2.8.6 Menu secret number
< sec. no. >
• Press Enter to confirm selection.
In the menu <sec.no.> you are able to select a code number to restrict
access to the parameter menus.
A number in the range between 00001 and 65535 can be chosen.
Once a sec.no. has been selected the following display will appear:
<- enter new se .. cret no.
>11111<
(0=none)
You can overwrite the number preset by the manufacturer.
• Press Enter to save the new number. A print-out of secret number as well
as serial number follows (printer mode: AUTO).
-----------------------------
Analyzer name
ser. no. C1770711 Sample print-out!
secret no=xxxxx x= number
-----------------------------
If zero is entered and saved instead of a number the secret number will not
be queried when a parameter menu has been selected.
The entry must be completed with Enter if a secret number shorter 5 digit is
stored.
Without entering a secret number the customer can:
- perform tests
- change the test method
without use of the secret number.
Once you have entered the secret number, you can
- perform calibration of tests
- perform settings and controls under UTILITIES and all its submenus.
2.2.8.7 Menu cuvette test
< cuvet. >
• Press Enter to confirm selection.