Operation
CoaDATA 501 – Operators Manual 2.1 Page 38
2.4.2 Error Messages
Error message
(Display) Cause Remedy
BREAK timeout
the maximum measuring time has been
exceeded
possibly no clotting; optical test
for clots; repeat test.
BREAK dark
preparation for measurement is too turbid dilute plasma or mix reagent.
BREAK top lim
exceeded measuring range (too high)
possibly caused by air bubbles
repeat test
BREAK bot lim
exceeded measuring range (too low)
repeat test
BREAK motor
mixer motor error occurred contact technical service
BREAK noise
loud noise after sample adjustment check for air bubbles or other
particles.
BREAK drift
measured curve drifted after reagent has
been added
check sample for air bubbles
break
measurement cancelled with Reset caused by user!
break jump
Measuring break because of a measuring
value jump (no clotting)
repeat test and if necessary
contact technical Service.
break readjust
message is displayed if the light value is too
dark during the adjustment phase
Repeat test and dilute plasma if
necessary.
mg/dl <2.0
converted value is lower than the
parameterized minimum value.
check analytical steps and the
conversion parameter.
mg/dl <200.0
converted value is bigger than the
parameterized maximum value.
see above (mg/dl <2.0)
Err div0
Err log0
Err over
deviation through 0 during conversion
comp. of the logarithm from a negat. Value
computation overflow.
check conversion parameter,
if necessary, contact technical
Service
SYSTEM FAILURE:
EPROM: Sxxxx S9999
EPROM check sum error
xxxx=set value; yyyy=actual value
contact technical Service
SYSTEM FAILURE:
MPU-RAM: Sxxxx
MPU-RAM on address Sxxxx in error
contact technical Service
SYSTEM FAILURE:
ext. RAM: Sxxxx
external RAM on address Sxxxx in error contact technical Service
parameter-error!
press any key ...
check sum error for parameters in EEPROM
contact technical Service
ERROR/ERROR rekursiv
software error
contact technical Service
All errors will cancel the current measurement.
NOTE