Operation
CoaDATA 501 – Operators Manual 2.1 Page 16
2.2 Method Parameterization
2.2.1 PT-parameterization
Method parameters in the analyzer have been preset by the manufacturer.
Prior to performing clotting analysis you must update the method parameter
for the reagent used.
Set analyzer to STANDBY mode
< 1 PT >
STANDBY
• Press Mode. The analyzer will request you to enter an up to 5 digit long
secret number (factory setting: 11111). For additional information refer to
chapter 2.2.8.6 Menu secret number.
<- secret no.:
..
_________
If the wrong number was entered STANDBY will be displayed. As soon as
the correct number has been entered, the following display will appear:
<1.conv>
Press the arrow key -> to display the following menus: <1. conversion>,
<2. conversion>, <replication>, <measurement> and <cuv remove
detec.>.
Overview PT parameterization:
<1.conversion>
<curve>
input of a 9-point calibration curve under or
<quick>
to enter the 100%-value and slope or
<none>
for no conversion
<2.conversion>
<INR>
input of the ISI-value for INR or
<ratio>
input of normal value for ratio calculation or
<none>
for no conversion
<replication>
select single or double determination
<single>
and the coefficient of variation
<double>
(CV 1-20%).
<measurement>
start reagent volume, reagent lot. no.
and sample incubation time.
<cuv remove detect.>
refer to chapter 3 Software
<ON>
activates the automatic cuvette detection
<OFF>
deactivates the automatic cuvette detection.
(default)