Operation
CoaDATA 501 – Operators Manual 2.1 Page 13
The results will be displayed consecutively for duration of 5 sec. The printer
will automatically print the results. The last message cuv out then press
"Reset" will request the removal of the cuvettes from the measuring
channels.
<- mean time =
..
12.4s Display of mean value in [sec.]
% = 91.0
Display of activity in [%]
INR = 1.05
Display of INR
If ratio is selected under 2nd conversion, ratio will be displayed instead of
INR.
<-cuv out, then
.. press „Reset“
Open the light protection cap.
Remove cuvette from measuring channel and confirm by pressing Reset-
key (see chapter 3 Software).
The analyzer is now ready for additional measurements.
cuv in 1
Continue as described for additional measurements.
The timer can be started or stopped manually by pressing the
Start-key.
Refer to function keys in chapter 1.2
2.1.4 How to change methods
Methods can only be changed from STANDBY.
< 1 PT >
STANDBY
• Press Esc to switch to STANDBY.
• Press the right arrow; the next method aPTT is displayed as ready-to-
measure.
• Press the arrow-key --> (arrow-key <-- back); the following methods and
the UTILITIES Menu will be displayed:
< 1 PT >
< 2 aPTT >
< 3 Fib. 1 >
g/l
< 4 Fib. 2 >
mg/dl
< 5 Thrmb >
< 6 Intr. >
< 7 Extr. >
Can be overwritten by ChipCARD!
< UTILIT >
NOTE