5.3 Technical Data
Instrument type Analyzer for determination of plasmic clotting.
Application coagulometric tests such as PT, aPTT, TZ,
Fibrinogen, single factors FII - FXII
Restrictions only for traditional, coagulometric clotting tests (no
chromogenic substrates).
Operation manual
Measuring principle turbodensitometric; opto-mechanical with
automatic zero adjustment and magnetic stir bar
for homogenizing of the test suspension and
increased sensitivity.
Sensitivity PT > 10% of norm
Test throughput PT 30/h, aPTT 15/h, +/- 10 tests/h
Cuvette volume min. 150 µl, max. 300 µl (test suspension)
Calibration manual input of calibration points, method
dependent
Software loaded in memory
Programmed PT, in sec, %, Ratio, INR (combinations)
methods aPTT, in sec, and Ratio
Fibrinogen, in sec, g/l,
Fibrinogen in sec, mg/dl
Thrombin T., in sec
Intr. Factor, in %
Extr. Factor, in %
Light source LED, light emitting diode
Display 1 lines with 8 characters, liquid crystal
Display
Processor 80552 (single chip microcontroller)
Incubation block controlled at 37.4°C +/- 0.3°C
Measuring channels 1
Light protection caps: for Thrombi-Tips
optional for yellow tips by Eppendorf
Reagent vials for 1 position, diameter 23.0 mm
Cuvette positions 4
Disposables cuvettes, paper for thermal printer, tips
Measuring timer max. 600 sec
Interfaces RS 232 C
(optional printer) ChipCARD reader