Operation
CoaDATA 501 – Operators Manual 2.1 Page 28
- <OFF> Cuvette detection occurs only when a cuvette is placed into
the measuring channel. The removal of cuvettes must be
confirmed by pressing the Reset-key.
• Select the desired cuvette detection.
• Press Enter-key to confirm entry.
<- select: ESC.
.. = work ENTER= more parameters
• Press Esc to access the measuring mode or Enter to add or verify
additional parameters
2.2.4 Fibrinogen 2 [mg/dl] - parameterization
<4 Fib.2> same as described under 2.2.3 Fibrinogen g/l <3Fib. 1>.
2.2.5 Thrombin time parameterization
<5Thrmb> same as described under 2.2.2 aPTT, but without 2nd
conversion possibility.
2.2.6 Intrinsic Factor parameterization
Intr. Factors <6 Intr. > (Factor VIII, Factor IX, Factor XI, or Factor XII) can
be parameterized with this menu. >intrinsic factors< = endogenous
clotting factors.
The following menus can be accessed under method parameters:
<1.conversion>
<curve>
input of a 9-point calibration curve under or
<none>
no conversion
<replication>
select single or double determination
<single>
and the coefficient of variation
<double>
(CV 1-20%).
<measurement>
start reagent volume, reagent lot. no.
and sample incubation time.
<cuv remove detect.>
refer to chapter 3 Software
<ON>
activates the automatic cuvette detection
<OFF>
deactivates the automatic cuvette detection.
(default)
2.2.7 Extrinsic Factor – parameterization
<1.conversion>
<curve>
input of a 9-point calibration curve under or
<none>
no conversion
<replication>
select single or double determination
and the coefficient of variation
<measurement>
(CV 1-20%).
start reagent volume, reagent lot. no.
<cuv remove detect.>
and sample incubation time.
<ON>
<OFF>
refer to chapter 3 Software
activates the automatic cuvette detection