2.2.2 aPTT - parameterization
• Change to STANDBY mode.
• Select method < 2 aPTT >
< 2 aPTT >
• Press Mode; and enter the secret no.
<- secret no.:
:_________
(Preset to 11111)
If you enter the wrong number STANDBY will appear.
The following dialog will appear as soon as the correct number has been
entered:
<2.conv>
The menus <2. conversion>, <replication>, <measurement> and <cuv
remove detec.> will be displayed by pressing the right arrow key.
Overview over aPTT parameterization
<2.conversion>
<ratio>
input of normal value for ratio calculation or
<none>
for no conversion
<replication>
select single or double determination
<single>
and the coefficient of variation
<double>
(CV 1-20%).
<measurement>
start reagent volume, reagent lot. no.
and sample incubation time.
<cuv remove detect.>
refer to chapter 3 Software
<ON>
activates the automatic cuvette detection
<OFF>
deactivates the automatic cuvette detection.
(default)
2nd conversion
<2.conv>
Selection Conversion
• Press Enter to confirm the selection.
ratio
< ratio >
• Press Enter to confirm the selection.
100% =
Example!