Labitec CoaDATA 501 - 2.2.3 Fibrinogen 1 [g;l] - parameterization

Labitec CoaDATA 501
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
CoaDATA 501 – Operators Manual 2.1 Page 23
27.5 s
Enter a normal value
• Press Enter to confirm the entry.
<- select: ESC.
.. = work ENTER= more parameters
• Press Esc to access the measuring mode or Enter to add or verify
additional parameters.
replication
< replic >
• Press Enter to confirm the selection.
< single >
or <double> for double determination
Only if double determination has been selected the following display for
entry of the coefficient of variation of the individual values will be displayed.
If this value is exceeded the message "mean error" will appear in the display
and print-out.
<- coef. variat
ion>
possible entry range
n = 5%
1% - 20%
Enter the coefficient of variation
• Press Enter to confirm the entry
<- select: ESC.
.. = work ENTER= more parameters
• Press Enter to add additional parameter or Esc to access the
measuring mode.
measurement
< measur >
• Press Enter to confirm the selection.
<- incubat.
=120 s
Enter the correct sample incubation time.
• Press Enter to confirm the selection.
A dialog will request the entry of start reagent volume and
reagent lot no.:
<-Start Reagent