Operation
CoaDATA 501 – Operators Manual 2.1 Page 26
5.28 g/l
Point of greatest concentration and
6.4 s
shortest clotting time!
• Enter the concentration and clotting time and confirm the entry with
Enter.
2.point:
2.53 g/l
11 s
• Enter the concentration and clotting time and confirm the entry with
Enter.
Proceed as described to enter additional points on the calibration curve
The following display appears once the last calibration curve point has been
confirmed:
<- select: ESC.
.. = work ENTER= more parameters
• Press Esc to access the measuring mode or Enter to add or verify
additional parameters.
replication
< replic >
• Press Enter.
< double >
Selection of double determination
• Press Enter to confirm the selection
Only if double determination has been selected the following display for
entry of the coefficient of variation of the individual values will be displayed.
If this value is exceeded the message "mean error" will appear in the display
and print-out.
<- coef. variat
ion>
Possible entry:
n = 10%
1% - 20%
Enter the coefficient of variation
• Press Enter to confirm the entry