EasyManua.ls Logo

LabJack U3 - Readmem (Readcal)

LabJack U3
120 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.2.6
Reads er
0x2A a (block numbers 0-7).
omm a which consists of 96 bytes
lock numbers 0-2). Do not call this function while streaming.
ReadMem (ReadCal)
1 block (32 bytes) from the non-volatile user or calibration memory. Command numb
ccesses the user memory area which consists of 256 bytes
and number 0x2D accesses the calibration memory are
C
(b
Command:
Byte
0 Checksum8
10xF8
20x01
3 0x2A (0x2D)
4 Checksum16 (LSB)
5 Checksum16 (MSB)
60x00
7BlockNum
ponse:
Res
Byte
0 Checksum8
10xF8
20x11
3 0x2A (0x2D)
4 Checksum16 (LSB)
5 Checksum16 (MSB)
6 Errorcode
70x00
8-39 32 Bytes of Data
95