EasyManua.ls Logo

LaMotte TC-3000 - Liquid Chlorine Calibration Setting; Setting Liquid Chlorine Calibration Value; Testing Tips for Chlorine

LaMotte TC-3000
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12. Open the meter lid. Insert the
sample into the chamber. Align the
index notch on the positioning ring
with the index arrow on the meter.
Close lid, select *IOK to select
Total Chlorine.
Free Chlorine (L)
1.00 ppm
* Total Chlorine
~
16:02:19 01/04/05
13. Record the result as Total Chlorine. Total Chlorine (L)
1.25 ppm
* Scan Sample ~
16:02:19 01/04/05
14. SubtracttheFreeChlorinereading
from the Total Chlorine reading to
obtain the concentration of
Combined Chlorine.
Tot al
Chlorine
Free
Chlorine
=
Combined
Chlorine
15. Press OFF to turn the meter off or
press to exit to a previous menu
or make another menu selection.
* W arning: Reagents marked with an * are considered to be potential health hazards.
To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD
or our web site. To obtain a printed copy, contact us by e-mail, phone or fax.
Note: For the most accurate results, samples over 6 ppm chlorine should be diluted
with chlorine demand free water and re-tested.
Note: The meter will remember the last scanned blank reading. It is not necessary to
scan a blank each time the test is performed. To use the previous blank reading,
instead of scanning a new one, scroll to Scan Sample and proceed. For the most
accurate results, the meter should be blanked before each test and the same tube with
tubepositioningringshouldbeusedfortheblankandthereactedsample.
CALIBRATION & ANALYSIS CHLORINE 57

Table of Contents

Other manuals for LaMotte TC-3000

Related product manuals