TOT AL CHLORINE
1. Press *IOK to turn the meter on.
1.3
2. Press *IOK to select Measure. Main Menu
* Measure
Data
Options
16:02:19 01/04/05
3. Scroll down and then press *IOK
to select Chlorine.
Measure
Turbidity
Color
* Chlorine
16:02:19 01/04/05
4. Press *IOK to select Test Free
Chlorine.
Chlorine (L)
Test Free
* Test Total
16:02:19 01/04/05
5. Rinse a clean tube (0290) with the
sample water. Fill the tube to the
10 mL line with the sample water.
Dry the tube with a lint-free cloth.
Put on a dry positioning ring. Cap
the tube.
6. Open the meter lid. Insert the tube
into the chamber. Align the index
notch on the positioning ring with
the index arrow on the meter. Close
the lid and press *IOK to select
Scan Blank.
Total Chlorine (L)
* Scan Blank
16:02:19 01/04/05
58 CALIBRATION & ANALYSIS • CHLORINE