EasyManua.ls Logo

Lamtec LT 2 - Page 67

Lamtec LT 2
105 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Commissioning / Shutdown
65
6.4.4 O
2
calibration with test gas
Connect test gas at the hose connector “test gas” of the PCB, with a prepressure of
0,3bar.
Example for test gas: 3 Vol.% O
2
in N
2
NOTE:
The calibration/check with O2- test gas only works with O2 values < 18%.
Menu function
When you press the [cal] menu key, “Start calibration” appears in the display.
Open the shut-off plug valve in the PCB
Set the pressure rise “dp” to 2...3mbar via the test gas throttle in the PCB
• Start calibration
When calibration starts, there is a waiting time of 5 seconds before the values are
displayed.
Change the O2-value with the cursor keys
• Press OK
Following successful calibration, close the shut-off plug valve
CAUTION:
If you are calibrating the measurement value using the display and operating unit
cal, you must confirm the new calibration value with “ENTER” or “OK”. Otherwise, the
menu ends automatically after 15 seconds, and the new calibration value is rejected.
NOTE:
If there is no possibility for a reference measurement or for tes gas, you have to enter
the probe temperature of the test report in parameter 141,to get a correct
measurement value.
But this method functions only with a new probe / sensor.
- O2 + O2

Table of Contents

Related product manuals