WARNING!
Malfunctions on the overtemperature protection or low-level
protection are not detected
Burns, scalding, fire
Check the Tmax function and low-level protection on a
regular basis.
WARNING!
Risk of refrigerant circuit bursting from excessive ambient tem-
peratures while device is inoperative
Impacts, cutting, device damage
Observe the permitted storage and operating tempera-
tures.
CAUTION!
Touching hot/cold equipment parts and accessories
Hot and cold burns
Never touch parts that are labeled with the warning symbol
“Hot”.
CAUTION!
Automatic device start with the timer
Scalding, cold burns, injury
Before using the timer, ensure that all preparatory meas-
ures for intended use have been implemented!
NOTICE!
Risk of refrigerant escaping from cooling water circuit
Device damage
Before decommissioning the device or if there is a risk
of freezing, drain the cooling water circuit of the cooling
unit using compressed air or an industrial vacuum cleaner
(watertight). Blow compressed air through the circuit.
The following information is only relevant for the Integral T device:
V6
Integral Process Thermostats and High-Temperature Thermostats 87 / 198