6300 Rev H User’s Manual
125-9106A Page 56
Row 2 right cell shows the height correction. This should normally be zero.
Rows 3 through 14 are the points of the profile. This profile has only 12 points. For each of these points
the columns contain the following information:
Column Information
A Pitot units
B Pitot mode
C Pitot target value
D Pitot Rate target value
E Static units
F Static mode
G Static target value
H Static Rate target value
In the sample profile, Row 3, which is the first point in the profile, tells the 6300 to go in Control mode to
100 knots at 150 knots/min on the Pitot side, and in Control mode to 0 feet at 6000 feet/min on the Static
side.
Row 4 (point 2) makes the 6300 go to 200 knots at 150 knots/min and 0 feet at 6000 ft/min.
Row 5 (point 3) puts the Pitot side into Leak mode while still keeping Static in Control at 0 feet.
Row 6 (point 4) puts the Pitot side back into Control and takes it down to 100 knots at 150 knots/min.
The Static side is simultaneously moved up to 1000 feet at 6000 ft/min.
Row 7 (point 5) takes Pitot up to 200 knots at 150 knots/min and Static up to 5000 feet at 6000 ft/min.
Row 8 (point 6) takes Pitot up to 300 knots at 150 knots/min and Static up to 10,000 feet at 6000 ft/min.
Row 9 (point 7) takes Pitot down to 100 knots at 150 knots/min and Static up to 18,000 feet at 6000
ft/min.
Row 10 (point 8) keeps Pitot at 100 knots and puts Static into Leak mode while at 18,000 feet.
Row 11 (point 9) keeps Pitot at 100 knots and ramps Static in Control mode to 30,000 feet at 6000 ft/min.
Row 12 (point 10) keeps Pitot at 100 knots and ramps Static to 50,000 feet at 6000 ft/min.
Row 13 (point 11) takes Pitot down to 20 knots at 150 knots/min and ramps Static back to “Ground” at
6000 ft/min.
Row 14 (point 12) puts both Pitot and Static into Measure mode.
Row 15, column A, indicates the end of the Profile.