13
Leica RM2245
3.3 Technical Data
General
Approvals: The instrument-specific marks are located next to the
identification label.
Nominal supply voltages: 100/120/230/240 V AC ±10 %
Nominal frequency: 50/60 Hz
Maximum power consumption: 70 VA
Protection class
: I
Power fuses: 2 x T 1.0 A UL listed
Pollution degree
: 2
Overvoltage category
: II
Maximum heat emission: 70 J/s
Operating temperature range: +10 °C to +35 °C
Temperature range during storage: +5 °C to +55 °C
Relative humidity: max. 80 % non-condensing
Storage humidity: < 80 %
according to IEC-1010, UL 3101, EN 61010
Dimensions and weight
Basic instrument
Width (including handwheel): 413 mm
Width (excluding handwheel): 300 mm
Depth (including waste tray): 618 mm
Height (total): 305 mm (with tray on the hood)
Working height (knife blade): 100 mm (measured from the base plate)
Working height (knife blade): 168 mm (measured from the table)
Weight (without accessories): approx. 37 kg
Control panel
Width: 94 mm
Depth: 164 mm
Height: 50 mm
Height (in inclined position): 81 mm
Weight (net): approx. 0.450 kg