Information for the tester
50
EMC (Immunity): IEC 61326-1:1997
IEC 61000-4-2 8 kV (air) Perf. Criteria B,
IEC 61000-4-3 3 V/m Perf. Criteria A,
IEC 61000-4-4 0,5 kV Perf. Criteria B
IEC 61000-4-5 1 kV Perf. Criteria B
IEC 61000-4-6 3 V Perf. Criteria A
IEC 61000-4-8 Level 4, 30A/m Perf. Criteria A
Quality system: developed, constructed and manufactured
complying to DIN ISO 9001
Auxiliary power: Six 1.5 V alkali-manganese-batteries (IEC LR6) or
1.5 V zinc-carbon-batteries (IEC R6) or
1.2 V nickel-cadmium-storage batteries
Battery life IEC LR6 : ≥ 3500 measurements
at max. load IEC R6 : ≥ 1200 measurements
(R
iso
- and R-measurement): NiCd : ≥ 1200 measurements
Case: Cycoloy, shockproof and scratchproof thermoplastic
material
Dimensions: 234 mm (W) x 175 mm (D) x 115 mm (H) incl. lid
and compartment for accessories.
Weight: approx. 1,5 kg w/o batteries and w/o accessories