Information for the tester
51
Measurement of Voltage
Measuring Range Display Range Resolution Frequency Range
1 – 400 V 0 – 500 V 1 V 46 – 65 Hz
Intrinsic error: ± (1 % of m.v. + 1 digit) for sinusoidal signal
Internal resistance: approx. 470 kΩ (U
L-N
: approx. 940 kΩ)
Measuring sequence: approx. 3 measurements/s
Permissible overload: U
rms
= 600 V AC
Measurement of frequency
Measuring Range Resolution Dynamic Range
15,3-99,9-457 Hz 0,1-1 Hz 1-400 V (U
L-N
: 4-400 V)
Intrinsic error: ± (0.1 % of m.v. + 1 digit)
Internal resistance: approx. 470 kΩ (U
L-N
: approx. 940kΩ)
Measuring sequence: approx. 3 measurements/s
Permissible overload: U
rms
= 600 V AC
Voltages outside the specified frequency range (e.g. direct voltage) cannot
be measured.
Fault Current Circuit Breaker Test (RCD)
With/without tripping, with/without selective (x2), with/without probe, x5 and ramp function for
determination of the tripping current
Measuring method: Lowering of voltage (without probe) or current-voltage
measurement (with probe) as per IEC 61557-6
Nominal voltage: 101 - 253 V sinusoidal
Frequency range: 46 - 65 Hz
Mains impedance angle:
1)
≤ 18 °; cos ϕ ≥ 0.95
Additional error
for cos ϕ < 0.95:
1)
+..% of m.v. = (1 - cos ϕ) x 100
Permissible overload: U
rms
= 600 V AC (measurement is not started)
Electrical safety is given during and functional safety after overload.