To get down to the minimum detected leak rate range some conditions
must be fulfilled:
PhoeniXL has to run at least 20 minutes
Ambient conditions must be stable (temperature, no vibration/accelera-
tions.)
The part under test has been evacuated long enough without using the
zero function (background is no longer decreasing)
ZERO must be active
1.3.3.2 Electrical Data
1.3.3.3 Other Data
1.3.3.4 Ambient Conditions
Power supply 230 V, 50 Hz
115 V, 60 Hz
Power consumption 350 VA
Type of protection IP40
Power cords (EU, USA, UK) 2.5 m
Valves solenoid
Dimensions (L × W × H) in mm 495 x 456 x 314
Weight in kg 35.5
Noise level dB (A) < 54
max. Audio alarm dB (A) 90
Contamination level (to IEC 60664-1) 2
Overvoltage category (to IEC 60664-1) II
For use within buildings
Permissible ambient temperature (during operation) +10 °C … +40 °C
Permissible storage temperature -10 °C … +60 °C
Max. rel. humidity 80% (up to 31°C)
linear decreasing to
50% at 40°C
Max. permissible height above sea level
(during operation)
2000 m