2.2 Electrical Connections
2.2.1 Mains Power
Generally the local regulations for electrical connections must be obser-
ved.
Caution Before connecting the PhoeniXL to the mains you must make
sure that the mains voltage rating of the PhoeniXL coincides with
the locally available mains voltage. The instrument must exclusi-
vely be connected to a single phase supply with fuses for instal-
lation (Circuit breaker 10A max. according to IEC/EN 60898 with
tripping characteristic B).
Attention Only 3-core mains cables having a protection ground conductor
must be used. Operation of the PhoeniXL where the ground
conductor has been left unconnected is not permissible. The
PhoeniXL can be damaged when using the wrong voltage. The
voltage must be in the range 230V (+/- 5%) or 115V (+/- 5%)
depending on the version.
The mains voltage rating for the PhoeniXL can be read off from the name
plate beneath the mains socket Fig. 3/4 at the side. This voltage is fixed
and can not be changed.
A separate fuse for each of the mains conductors has been integrated
into the mains switch.
The mains voltage is applied to the instrument via the detachable mains
cable which is supplied with the instrument. A mains socket Fig. 3/4 is
available for this purpose at the side of the instrument.