5.4.1 Call for Calibration
In all modes the Soft Key no. 5 is used to get to the calibration routine.
Refer to Chapter 7 Calibration for further information regarding calibration.
5.4.2 Speaker Volume
On the left hand side two loud speaker symbols are shown, combined
with the signs + and - (Fig. 14 & 15). By pressing the corresponding soft-
keys (Soft Keys no. 2 and no. 3) the volume can be adjusted for conveni-
ent loudness. In the bottom line of the display another loud speaker
symbol is shown, combined with a number. This number indicates the
level of the current loudness (ranges from 0 to 15).
Refer to Chapter 6.4.5 Volume for information on loudness, alarms, and
sound tracks.
5.4.3 Status Line in the Display
The status line at the top of the display (Fig. 14 and Fig. 15) informs about
(reading from left to right):
Fig. 14 Display: Measurement Mode with bargraph
Symbol of display Meaning
Explanation
Volume level
Please refer to Chapter 5.4.2
Speaker Volume.
S1 Trigger 1
If the trigger values are excee-
ded these signs are inverted.
(White on black background.)
S2 Trigger 2
see: Trigger 1
S3 Trigger 3
see: Trigger 1
! Warning triangle
Please refer to Chapter 8.1
VAC Working mode
VAC or SNIFF indicate which
working mode was selected
FINE Vacuum area
Depending on the inlet pressure
the Phoenix may be in GROSS,
PRECISION (PhoeniXL
300 dry
only) or FINE mode, which is
indicated here (Chapter 4.1.3 )