Head
GA10218_0101 - 5/2002
⏐ 72
6.6.4.3 RS232
Main menu > Settings > Interfaces > RS232
The calibration function of Leak Ware is not suitable with the use of a
Phoenix.
6.6.4.4 Define PLC outputs
Main menu > Settings > Interfaces > Define PLC outputs
The following relay outputs are available for further signal processing. The
maximum rating for the relay contacts is 60V AC/1A.
The contacts are numbered from bottom to top.
The fundamental output voltage is scaled logarithmic. The voltage output
ranges from 1 to 10 V with adjustable steps of 0,5 to 10 V per decade
(Fig. 26)
Fig. 26 Example of range of leak rate, log. 0.5 V/decade
Softkey 2: Baudrate and end sign
Settings for the baudrate selectable between 1200 and 19200, as well as
endsign between CR+LF, CR or LF.
Softkey 3: Data, Parity, Stop bits
The settings for Data ( 7 or 8 ), Parity ( Even, Odd, None ) and Stop bits (
1 or 2 ) can be selected
Softkey 7: RS 232 Protocol
The protocol from the RS 232 can be choosen between: L200 size, Dia-
gnostic, L200 Leakware and ASCII code.
Pin
Assignment
1 PLC in free selectable
2 PLC in free selectable
3 PLC in free selectable
4
GND
5 to 7 Digital out free selectable, 5 center contact, 6 normally open
contact, 7 normally closed contact
8 to 10 Digital out free selectable
11 to 13 Digital out free selectable
14 to 16 Digital out free selectable
012345678910V
-3 -1 +1 +3
10 10 10 10