Head
16
⏐
GA10218_0101 - 5/2002
1.3.4.3 Other Data
1.3.4.4 Ambient Conditions
1.4 Unpacking
Unpack the PhoeniXL immediately after delivery, even if it will be installed
later on.
Examine the shipping container for any external damage. Completely
remove the packaging materials.
Check the PhoeniXL is complete (see Chapter 1.4.1 Supplied Equipment)
and carefully examine the PhoeniXL visually.
If any damage is discovered, report it immediately to the forwarding agent
and insurer. If the damaged part has to be replaced, please contact the
orders department.
Retain the packaging materials in the case of complaints about damage.
1.4.1 Supplied Equipment
Leak Detector PhoeniXL
Set of fuses
Power cord
Folder with documents
– Operating unstructions PhoeniXL
– Spare Parts List PhoeniXL
2 L-type screwed connections (hose connections)
1 hose nozzle
Valves solenoid
Dimensions (L × W × H) in mm 495 x 456 x 314
Weight in kg 29.5
Noise level dB (A) < 54
max. Audio alarm dB (A) 90
Contamination level (to IEC 60664-1) 2
Overvoltage category (to IEC 60664-1) II
For use within buildings
Permissible ambient temperature (during operation)+10 °C … +40 °C
Permissible storage temperature -10 °C … +60 °C
Max. rel. humidity 80% (up to 31°C)
linear decreasing to
50% at 40°C
Max. permissible height above sea level
(during operation)
2000 m