E69 TMP error: Runup timeout Max. time after which the pump must enter its normal operation
mode has been exceeded.
Forevacuum or high-vacuum pressure to high: see chapter 6.6.7.1
to reduce the inlet pressure of the PhoeniXL
Bearing defective: Inform Leybold service for repair
E70 TMP error: TMP motor tempe-
rature too high (>90°C)
Max. motor temperature has been exceeded.
Forevacuum or high-vacuum pressure to high: see chapter 6.6.7.1
to reduce the inlet pressure of the PhoeniXL
Fan defective: Replace the fan
Ambient temperature too high: Feed cooler air to the PhoeniXL
Bearing defective: Inform Leybold service for repair
E71 TMP error: TMP could not be
indentified
Pump could not be identified or no pump is connected.
Pump not connected to Turbo.Drive S: Check connecting cable
E73 Emission off (p2 too high) The emission is switched off as soon as the pressure P2 > 0.2 mbar
or 1,5 mbar in the measurement mode. If after closing the inlet valve
the pressure drops again, the leak detector will revert to the standby
mode.
Air incrush in the measurement mode.
W76 Maximum of evacuation time
was exeeded.
Within the pre set evacuation time the pressure threshold of 100
mbar was not reached.
Test sample has got a GROSS leak.
Wrong setting ( too short) of the max. time of evacuation.
W77 Peak not in Range The signal maximum has shifted to mass range alignment limits.
Signal of leak rate was instable during mass adjustment. Calibrate
again.
Check the basic setting for the anode voltage through the service
menu.
Check calibrated leak.
W78 Differences of signal between
test leak open and closed is
too low.
The amplifier voltage difference between opened and closed calibra-
ted leak is less than 10 mV.
Calibrated leak has not been closed properly.
W79 Signal of test leak is too small Calibrated leak is too small or has not been opened. Preamplifier
voltage < 10 mV.
W80 Please calibrate machine
newly
The automatic request of calibration is activated (7.2.1) and has full-
filled at least one of the conditions:
30 minutes are passed since power on.
Temperature of the pre-amplifier has changed more than 5°C since
the last calibration.
Massadjustments were changed.
No. Displayed Message Description and possible solutions