Operation
59
300554859_002_C0 - 11/2016 - © Leybold
No. Designation Description
125 Actual bearing temp Calculated temperature of the bearing.
126 Bearing temp warning threshold Exceeding the bearing temperature warning threshold results in a warning.
128 Motor temp lower warning
threshold
Falling below the motor temperature lower warning threshold results in a war-
ning.
131 Motor temp lower error threshold Falling below the motor temperature lower error threshold results in an error
message.
132 Bearing temp error threshold Exceeding the bearing temperature error threshold results in an error message.
133 Motor temp error threshold Exceeding the motor temperature error threshold results in an error message.
140 Intermediate circuit current Mean value measurement of the intermediate circuit current. Corresponds to the
current consumption of the frequency converter.
150 Standby Frequency Standby operation frequency setpoint
171
[0..253]
Error Code Memory Indexed parameter for storing the most recent 40 error codes. The individual
error memory entries are accessed via this parameter with additional index
number. The last error code is accessed with index 0 and the oldest with index
39.
174
[0..253]
Hz @ Time of Error Access analogously as for parameter 171
176
[0..253]
Hours @ Time of Error Access analogously as for parameter 171
184 Converter Operating Hours Counts the operating hours of the converter during active pump operation.
249 Generator mode P249 = 0 : no return feeding in to the DC supply
P249 = 1 : return feeding in to the DC supply
Notice: take note of the maximum generator power of as otherwise the electro-
nics may suffer damage.
312
[0..17]
Converter part number Converter part number. One ASCII character per index.
315
[1..17]
Converter serial number
355
[0..17]
Pump serial number not used