Get information
l
Command: 0I! (the number zero, the capital letter I for information, ending
with an exclamation point).
l
Reply: 014LI-COR LI-7101.0ET100033 (identifying information for the
sensor, including the device address, manufacturer, model, firmware version, and
serial number).
Read current data from a group
The command to read data.
l
Command: 0R0!
l
Reply: 0+0.00+0.02+0.02+1.39+97.55+22.47+48.87+0
l
Command: 0R1!
l
Reply: 1+0.00+0.02+0.02+97.55+22.47+48.87+600+0
The two examples show data group 0 and data group 1. Enter 2 or 3 in place of the
0 or 1 to read those data groups.
Simple program with comments
The program in Listing 5-1 below is for a Campbell Scientific CR1000X, CR1000,
CR5000, or CR6 Datalogger. This example includes comments to show you which
parameters can be modified.
Listing 5-1. A program for Campbell Scientific, Inc. CR1000X, CR1000, CR5000, or CR6
datalogger.
'LI-710 simple logger
'Rev: 0.1
'Date: May 23, 2023
'Author: licor.com
Public LI710A(9) 'names an array and specifies that it has 9
variables
Alias LI710A(1)=et_l 'the first of 9
Units et_l=mm
Alias LI710A(2)=le_l
Units le_l=W/M^2
Alias LI710A(3)=h_l
Units h_l=W/M^2
Section 5. Connecting to Campbell Scientific dataloggers