Electrical Safety
DANGER! ELECTRICAL SHOCK HAZARD. Severe electrical shock can
result from operating the EVOS
®
FL Imaging System without its instrument
panels in place. Do not remove instrument panels. High
voltage contacts
are exposed when instrument panels are removed from the instrument.
Fuses
WARNING! FIRE HAZARD. For continued protection against the risk of
fire, replace fuses only with fuses of the type and rating specified for the
instrument.
Power
DANGER! ELECTRICAL HAZARD. Grounding circuit continuity is vital
for the safe operation of equipment. Never operate equipment with the
grounding conductor disconnected.
DANGER! ELECTRICAL HAZARD. Use properly configured and
approved line cords for the voltage supply in your facility.
DANGER! ELECTRICAL HAZARD. Plug the system into a properly
grounded receptacle with adequate current capacity.
Overvoltage Rating
The EVOS
®
FL Imaging System has an installation (overvoltage) category of II, and
is classified as portable equipment.