ApexRp Operators Manual
248083480-1 Rev 3 56
Register
Data Type
Description
40021
ASCII string
Model Name char[12], char [13]
40022
unsigned integer
Model Name char[14], char [15]
40023
unsigned integer
Flow Rate. See registers 40041-40042 for flow rate units.
Liquid Particle Counters and Samplers: Value equals flow
rate. For example: 100d=100. All Other Instruments:
Divide by 100 to get flow rate. For example: 100d=1.00
40024
unsigned integer
Record Count. Total number of records stored in the
counter.
40025
unsigned integer
Record Index. Zero based index to data in 3xxxx register
series. Must be lower than the record count (register
40024). Set this index to expose a counter’s record in the
3xxxx registers. Set to -1 to retrieve last record stored in
the counter.
40026
unsigned integer
Location number
Particle Counters: Specifies location of Particle Counter
where data was recorded. Must be 1 to 200 (maps to
location names associated with registers 40200 - 40999).
40027
signed integer
Real Time Clock (RTC) [high]. Displays instrument’s real-
time clock. Works in conjunction with 40028. Displays
date and time, in number of seconds since midnight,
1/1/1970. Can be generated by ANSI C/C++ time()
function.
40028
signed integer
Real Time Clock [low]
40029
unsigned integer
Initial Delay [high]. Works in conjunction with 40030.
Number of seconds to wait before starting the first
sample. Max value is 359,999, which equals 99h 59m 59s.
40030
unsigned integer
Initial Delay [low]
40031
unsigned integer
Hold Time [high]. Works in conjunction with 40032.
Number of seconds to wait between sample periods. Max
value is 359,999, which equals 99h 59m 59s
40032
unsigned integer
Hold Time [low]
40033
unsigned integer
Sample Time [high]. Works in conjunction with 40034.
Number of seconds to sample. Max value is 86,399,
which equals 23h 59m 59s.
40034
unsigned integer
Sample Time [low]
Table A-3 Sensor Settings Registers

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lighthouse Worldwide Solutions ApexR3p and is the answer not in the manual?

Lighthouse Worldwide Solutions ApexR3p Specifications

General IconGeneral
BrandLighthouse Worldwide Solutions
ModelApexR3p
CategoryRemote Control
LanguageEnglish