ApexRp Operators Manual
57 248083480-1 Rev 3
Data Set [high]. Works in conjunction with 40036.
Updates the instrument’s real time clock. Setting is the
number of seconds since midnight, 1/1/1970. This
number can be generated by the ANSI C/C++ time()
function.
Flow Unit - Defines the Unit that Flow Rate value is based
on. char[0], char[1]. (NULL-terminated string)
Flow Unit. char[2], char[3]
Calibration Reference Voltage (millivolts)
ApexRp: Calibration Due Date [high]. Indicates when
instrument is due for calibration. This number can be
generated by the ANSI C/C++ time() function.
Calibration Due Date [low].
Location ID[high]. Value is 0 when no location bracket is
present.
Location ID[low]. Value is 0 when no location bracket is
present. Value matches 40026 location number when
bracket is present.
Table A-3 Sensor Settings Registers