Page #59 background imageLoading...
Page #59 background image
ApexRp Operators Manual
248083480-1 Rev 3 58
Register
Data Type
Description
40060
signed integer
Last Sample Timestamp [high] (# of seconds since
midnight, 1/1/1970.).
40061
signed integer
Last Sample Timestamp [low].
40062
signed integer
Last Setting Change Timestamp [high] (# of seconds
since midnight, 1/1/1970.). Value indicates.
40063
signed integer
Last Setting Change Timestamp [low].
40064
signed integer
Run-time particle channel alarm high flags (bit 0 =
channel 1, …).
40065
signed integer
Run-time particle channel alarm low flags (bit 0 =
channel 1,…).
40066
signed integer
Run-time analog channel alarm high flags (bit 0 =
channel 1,…).
40067
signed integer
Run-time analog channel alarm low flags (bit 0 =
channel 1,…).
40068
unsigned integer
Software controlled RGB LED red channel. Uses
values from 0-100 for duty cycle percentage every
even second from UNIX time.
40069
unsigned integer
Software controlled RGB LED green channel. Uses
values from 0-100 for duty cycle percentage every
even second from UNIX time.
40070
unsigned integer
Software controlled RGB LED blue channel. Uses
values from 0-100 for duty cycle percentage every
even second from UNIX time.
40071
unsigned integer
Software controlled RGB LED red channel. Uses
values from 0-100 for duty cycle percentage every
odd second from UNIX time.
40072
unsigned integer
Software controlled RGB LED green channel. Uses
values from 0-100 for duty cycle percentage every
odd second from UNIX time.
40073
unsigned integer
Software controlled RGB LED blue channel. Uses
values from 0-100 for duty cycle percentage every
odd second from UNIX time.
40074
signed integer
ApexRp: Last Calibration Date [high]. Indicates when
instrument was last calibrated. This number can be
generated by ANSI C/C++ time() function.
40075
signed integer
ApexRp: Last Calibration Date [low]
Table A-3 Sensor Settings Registers

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lighthouse Worldwide Solutions ApexR3p and is the answer not in the manual?

Lighthouse Worldwide Solutions ApexR3p Specifications

General IconGeneral
BrandLighthouse Worldwide Solutions
ModelApexR3p
CategoryRemote Control
LanguageEnglish