ApexRp Operators Manual
248083480-1 Rev 3 68
Service LASER Status
0 = LASER is good. 1 - LASER needs to be serviced.
Bad Flow Status
0 = Flow rate is good. 1 = Flow rate is bad.
Particle Overflow Status
0 = No overflow 1 = Overflow occurred.
Instrument Service Status
0 = Instrument is working correctly.
1 = Service light is on. Instrument malfunction detected.
Threshold High Status
0 = Threshold not exceeded. 1 = Threshold exceeded.
Threshold Low Status
0 = Threshold not exceeded. 1 = Threshold exceeded.
Instrument Sampler Status
0 = Nominal Operation. 1 = Sampler Error.
ApexRp:
0 = LASER power in spec. 1 = LASER power out of spec.
ApexRp: LASER current status
0 = LASER current in spec. 1 = LASER power out of spec.
ApexRp: LASER supply status 0 = LASER supply in spec. 1 =
LASER supply out of spec.
ApexRp: LASER life status 0 = LASER life in spec. 1 = LASER
life out of spec.
ApexRp: No flow status
0 = There is flow in the unit. 1 = no flow in unit.
ApexRp: Photoamp supply status 0 = Photoamp supply in
spec. 1 = Photoamp supply out of spec.
Background status 0 = Photoamp background good. 1 =
Photoamp background out of spec.
ApexRp: Photodiode health status
0 = Photodiode good. 1 = Photodiode failure
ApexRp: Validation mode status
0 = Normal operation data. 1 = Validation dummy data.
ApexRp: Calibration due date status
0 = Unit has not passed calibration date. 1 = Unit has
passed calibration.
Table A-14: Instrument Current Status