D9S001ES1-0101
31
The gravity constant can be set according to the geographical zone of the country in the range 9761-9839 or 1 - 8.
Values according to the geographical zone:
Obr. 32 Gravitational constant depending on the geographical zone
Zone Latitude Gravitational constant [m/s²]
1 0 až 5 ° 9.780
2 5 až 15 ° 9.782
3 15 až 25 ° 9.786
4 25 až 35 ° 9.793
5 35 až 45 ° 9.802
6 45 až 55 ° default settings 9.810
7 55 až 65 ° 9.819
8 65 až 75 ° 9.825