4.1.2 Overview of error codes
Letter indicating the type of error Type of error
A Keyboard errors
B Scale system hardware errors
F Weighing system errors
1 2 3 4 5 6
Keyboard error code
(letter A)
Meaning
0x1 Long press (≥ 60) detected on the ZERO/T button (button No. 1)
0x2 Long press (≥ 60) detected on the WEIGHT/CLEAR button (button No. 2)
0x4 Long press (≥ 60) detected on the kg lb button (button No. 3)
0x8 Long press (≥ 60) detected on the Bed Exit Alarm ON/OFF button (button No. 4)
0x10 Long press (≥ 60) detected on the Bed Exit Alarm pause button (button no. 5)
0x20 Long press (≥ 60) detected on the Bed Exit Alarm mode selection button (button No. 6)
0x100 Error detected on ZERO/T button (button No. 1)
0x200 Error detected on WEIGHT/CLEAR button (button No.2)
0x400 Error detected on kg/lb button (button No. 3)
0x800 Error detected on the Bed Exit Alarm ON/OFF button (button No. 4)
0x1000 Error detected on the Bed Exit Alarm pause button (button No. 5)
0x2000 Error detected on the Bed Exit Alarm mode selection button (button No. 6)
Scale system hardware error code
(letter B)
Meaning
0x1 Overvoltage on the scale system processor
0x2 Undervoltage on the scale system processor
0x4 Voltage measurement error on the scale system processor
0x8 Overheating of the scale system processor
0x10 The processor temperature of the scale system is below the permitted limit
0x20 Error of temperature measurement of the scale module processor
0x80 Other hardware error
0x2000 Error of Bed Exit Alarm function
0x4000 Button measurement error (control measurement mismatch)