EasyManua.ls Logo

LINET Sprint 100

LINET Sprint 100
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D9S001ES1-010130
5.1.6 Gravitational constant setting
The gravitational constant can be set separately in the G CONST menu or during the scales calibration. The menu closes after 60
seconds of inactivity (the wheel in the upper right corner), any press resets the counter.
To access the CALIBRATION CONSTANT mode from the unlocked main menu (to unlock the menu, see chapter 5.1):
Select G CONST using the PLAY/STOP or PAUSE button
Confirm your selection by pressing WEIGHT/CLEAR
Select SET G by pressing the WEIGHT/CLEAR button
The „On screen keyboard“ screen opens
Select from the digits 0 - 9 to enter the value of the gravitational constant valid for the country where the bed is used
(without a decimal point, ie 9780 - 9825) or the value for the relevant geographical zone 1-8 (see Fig. 31 with the table).
The selection is made with the PLAY/STOP (scroll left) or PAUSE (scroll right) buttons. Confirm each digit with the WEIGHT/
CLEAR button. The character is written in the box
The character in the box can be deleted by pressing the ZERO/T button (if no value is entered in the box, the screen will be
left without confirming the entry)
After entering the constant value, move to the enter character and press the WEIGHT/CLEAR button
This confirms the value of the gravity constant - the value G is selected and at the SET G item is displayed OK

Other manuals for LINET Sprint 100

Related product manuals