EasyManua.ls Logo

LMI Selcom SLS 5000 - Warning Labels

Default Icon
62 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
03-09-05 19
necessary to be able to calculate the NOHD are MPE, max average power, beam
divergence, diameter of emergent laser beam (a) and Non Gaussian correction
factor. These parameters can be found as follows.
Parameter
Max average power Q-record
Beam divergence angle See section “ Sensor configuration”
Non Gaussian correction factor LMI use 2.5.
MPE See table above
Diameter of emergent laser beam (a) 5 mm
The NOHD value for each sensor can be found in the Q-record.
Warning labels
Products complies with IEC 60825-1 and with FDA performance standards for
laser products except for deviations pursuant to Laser Notice No. 50, dated
July 26, 2001.
The following labels are attached to the sensor:-All languages:
Figure 11: Manufacturers serial number label
Class 2

Table of Contents