Lode BV Excalibur Sport
73
D.3 Abort codes BP system
Prior.
Error message
Display
Possible error cause
1
ΑCHECK ECG
SIGNAL≅
0 / 0 mmHg
! ECG trigger not connected.
! ECG trigger properties not correct.
! ECG trigger polarity not correct.
! QRS pulse width smaller than 10 ms.
2
ΑCUFF
OVERPRESSURE≅
0 / 0 mmHg
! Strong arm movements during the blood
pressure measurement.
! Air hose is bent or blocked.
3
ΑAIR LEAK≅
0 / 0 mmHg
! Cuff is loose or not connected.
! Air hose or air-bubble leak.
! Air hose not connected.
4
ΑSERVICE
REQUIRED≅
BP
value
! Fatal error ...Get in touch with the company
Lode.
5
ΑCHECK ECG/
MICROPHONE≅
BP
value
! Poor patient hook-up (irregular ECG-signal)
or poor trigger signal.
AND
! Abnormal microphone-signal or damaged
microphone.
6
ΑCHECK
MICROPHONE≅
BP
value
! Wrong Microphone placement or defect
microphone.
7
ΑCHECK ECG/
ARTIFACT≅
BP
value
! ECG trigger signal failed during blood
pressure measurement.
Bad or irregular ECG-signal during the blood
pressure measurement.
AND
! Excessive movements of the patient during the
blood pressure measurement.
8
ΑARTIFACT/ARM
MOVEMENT≅
BP
value
! Excessive movements of the patient during the
blood pressure measurement.
9
ΑCUFF INFLATION
TOO LOW≅
BP
value
! Maximum cuff pressure adjusted too low.
! Inflation pressure adjusted too low. Repeat
measurement.
10
None
BP
value
D.4 Cleaning and disinfection of the cuff