EasyManua.ls Logo

Lode 925900 - BP System Error Codes and Cleaning; D.3 Abort Codes BP System

Lode 925900
89 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Lode BV Excalibur Sport
73
D.3 Abort codes BP system
Prior.
Error message
Display
Possible error cause
1
ΑCHECK ECG
SIGNAL
0 / 0 mmHg
! ECG trigger not connected.
! ECG trigger properties not correct.
! ECG trigger polarity not correct.
! QRS pulse width smaller than 10 ms.
2
ΑCUFF
OVERPRESSURE
0 / 0 mmHg
! Strong arm movements during the blood
pressure measurement.
! Air hose is bent or blocked.
3
ΑAIR LEAK
0 / 0 mmHg
! Cuff is loose or not connected.
! Air hose or air-bubble leak.
! Air hose not connected.
4
ΑSERVICE
REQUIRED
BP
value
! Fatal error ...Get in touch with the company
Lode.
5
ΑCHECK ECG/
MICROPHONE
BP
value
! Poor patient hook-up (irregular ECG-signal)
or poor trigger signal.
AND
! Abnormal microphone-signal or damaged
microphone.
6
ΑCHECK
MICROPHONE
BP
value
! Wrong Microphone placement or defect
microphone.
7
ΑCHECK ECG/
ARTIFACT
BP
value
! ECG trigger signal failed during blood
pressure measurement.
Bad or irregular ECG-signal during the blood
pressure measurement.
AND
! Excessive movements of the patient during the
blood pressure measurement.
8
ΑARTIFACT/ARM
MOVEMENT
BP
value
! Excessive movements of the patient during the
blood pressure measurement.
9
ΑCUFF INFLATION
TOO LOW
BP
value
! Maximum cuff pressure adjusted too low.
! Inflation pressure adjusted too low. Repeat
measurement.
10
None
BP
value
D.4 Cleaning and disinfection of the cuff

Table of Contents

Related product manuals