Only after confirmation with save the new setting is stored. With this option an ECG
apparatus wich offers P1 can be connected to the ergometer.
Note: In the original protocol the patient weigth and several PWC (Physical - Work Capacity)
values must be entered. At the end of the protocol these PWC values where send to the ECG
apparatus. This P1 protocol does not offer PWC calculation. The advantage of this
implementation is that any programmed protocol can be executed when using a P1 supported
ECG apparatus. PWC's send to the ECG are all filled out with 0.
B.3.7 Operation from basic control panel
With the basic control panel as described in the standard user guide you had the possibility to
switch your KEYBOARD ON or OFF. With the programmable control panel the basic (test
subject) side of the unit is default OFF. It is possible to control the ergometer with the buttons
of the basic control panel instead of the programmable control panel. You can use this option
for service purposes or to program the display of the basic control panel.
During the start up of the ergometer you can press one of the buttons at the basic side to
switch operation to this basic control panel, you will then loose the control function of the
programmable side. To go back to the default control situation you have to restart the
ergometer without pressing any button.
The TEST FUNCTIONS are only available at the basic control panel. They are meant for
service personnel only. Contact your supplier in case service of the device is required. If,
accidentally, you get into to the TEST FUNCTIONS it can cause damage to the
ergometer operation and guarantee will be lost!
B.3.8
Protocol
In this paragraph the program possibilities under the menu PROTOCOL will be described:
You can select the menus PROTOCOL EDIT; PROTOCOL RUN and PROTOCOL
DELETE.
B.3.8.1 Protocol edit
The menu PROTOCOL EDIT allows you to program new exercise protocols or edit an
existing protocol. The memory capacity will allow you to save 50 protocols, all existing of
maximal 15 steps each. Pressing this menu will show you the list of protocols which have
already been made and the menu New.
New
name1
name2
esc ok
run protocol
EDIT PROTOCOL
Delete protocol
prot edit