Lode BV Excalibur Sport
79
E.3 Using the SpO
2
Option
The SpO
2
is come together with the P-Display. After startup the ergometer indicates correct
operation of the SpO
2
device by showing the following screen for a few seconds:
When in the sub menu SET DISPLAY of the SYSTEM PARAMETER SpO2 is selected to be
displayed. As follows,
The following can be displayed.
The SpO2 option also measures the heart rate and this value is displayed when this display
setting is selected. The heart rate value measured by the SpO
2
option can not serve as a trigger
source for the blood pressure option.
The SpO2 option also report errors due to bad signals or improper handling. The following is
displayed when the sensor is not connected:
The following errors can be displayed:
SNSD Sensor Disconnect – An absence of signal, meaning the sensor is gone
OOT Out Of Track – An absence of consecutive good pulse signals
LODE
SpO2 Initialized
TERMINAL
P= 0W 0 rpm
= 55/min SpO2= 98
ecs esc
SpO2
On
Esc Off On next
TERMINAL
P= 0W 0 rpm
= 55/min SpO2= SNSD
ecs esc