p/n 18603-001, Rev. C LTV
®
1200/1150 Ventilator Service Manual Page 8-77
Internal Battery Pack
Parts required for replacement: Tools required:
• LTV
®
1200/1150 Internal Replacement
Battery
P/N 18608-001.
• Torque driver (120 in-oz/0.84 Nm)
• 3/32” Hex bit
CAUTION
Internal Battery Use - The internal battery is intended for use in LTV
®
ventilators during
short periods while switching between external power supply connections, emergency
situations or short duration transports. The length of time the ventilator will operate on
battery power is a function of many factors such as settings, charge level and condition or
age of the battery; therefore, the use of the internal battery as a standard operating practice
is not recommended.
To remove and replace the battery pack:
1) Position the LTV
®
1200/1150 ventilator as shown (front side down) on a clean, dry
work surface and use a 3/32” hex bit to remove both button socket head cap screws
(2, p/n 18296-001) from the Battery Cover.
2) Open the cover and take note of how the battery connector wires are positioned
between the side of the case and the battery.