Page 9-6 LTV
®
1200/1150 Ventilator Service Manual p/n 18603-001, Rev. C
D) Return the High Pres. Limit
setting to 100 and clear all
alarms.
E) Turn the ventilator off, leaving the
audible alarm sounding and
measure the plug resistance.
Resistance >1.0 mega ohms
F) Press the Silence Reset button
to clear the audible alarm and
measure the plug resistance.
Resistance 2.3 ohm
G) Disconnect the Patient Assist
Cable, Normally Closed (P/N
10779) and connect a Patient
Assist Cable, Normally Open
(P/N 10780) to the ventilator
Patient Assist port.
H) Turn the ventilator back on in
breathing mode, clear all alarms
and measure the plug resistance.
Resistance >1.0 mega ohms
I) Create a High Pressure alarm by
changing the High Pres. Limit
setting to 5 and measure the plug
resistance.
Resistance 2.3 ohm
68
J) Return the High Pres. Limit
setting to 100 and clear all
alarms.
K) Turn the ventilator off, leaving the
audible alarm sounding and
measure the plug resistance.
Resistance 2.3 ohm
L) Press the Silence Reset button
to clear the audible alarm and
measure the plug resistance.
Resistance >1.0 mega ohms.
7) Check date, time and usage by operating the ventilator using the test lung for 1 hour.
A) Set the date to the current date. Note the current date and date format.
B) Set the time to 23:15.
C) Note the current usage meter.
D) Run the ventilator connected to a
test lung for 1 hour.
Verify that no alarms occur.
E) Power the vent off and on. Check
the date.
Date has incremented correctly.
F) Check the time. Time has incremented correctly.
68
This measurement can best be obtained using the relative (delta) measurement of a DMM.