Page 7-6 LTV
®
1200/1150 Ventilator Service Manual p/n 18603-001, Rev. C
Symptoms Possible Causes What to Do
High f OFF is
displayed.
The High Breath Rate
alarm is turned off.
This is an informational message only
(See the LTV
®
1200 or the LTV
®
1150
Ventilator Operator's Manual for an
explanation of this feature).
HI f/Vt OFF is
displayed.
The SBT High f/Vt alarm is
turned off during the SBT
mode of ventilation.
This is an informational message only
(See the LTV
®
1200 or the LTV
®
1150
Ventilator Operator's Manual for an
explanation of this feature).
LO f/Vt OFF is
displayed.
The SBT Low f/Vt alarm is
turned off during the SBT
mode of ventilation.
This is an informational message only
(See the LTV
®
1200 or the LTV
®
1150
Ventilator Operator's Manual for an
explanation of this feature).
LO PEEP OFF is
displayed.
The Low PEEP alarm is
turned off.
This is an informational message only
(See the LTV
®
1200 or the LTV
®
1150
Ventilator Operator's Manual for an
explanation of this feature).
SBT f OFF is
displayed.
The SBT High Breath Rate
and SBT Low Breath Rate
alarms are turned off
during the SBT mode of
ventilation.
This is an informational message only
(See the LTV
®
1200 or the LTV
®
1150
Ventilator Operator's Manual for an
explanation of this feature).
SBT f/Vt OFF is
displayed.
The SBT High f/Vt and
SBT Low f/Vt alarms are
turned off during the SBT
mode of ventilation.
This is an informational message only
(See the LTV
®
1200 or the LTV
®
1150
Ventilator Operator's Manual for an
explanation of this feature).
SBT HI OFF is
displayed.
The SBT High Breath Rate
alarm is turned off during
the SBT mode of
ventilation.
This is an informational message only
(See the LTV
®
1200 or the LTV
®
1150
Ventilator Operator's Manual for an
explanation of this feature).
SBT LO f OFF is
displayed.
The SBT Low Breath Rate
alarm is turned off during
the SBT mode of
ventilation.
This is an informational message only
(See the LTV
®
1200 or the LTV
®
1150
Ventilator Operator's Manual for an
explanation of this feature).
SBT TIME is
displayed.
The SBT mode of
ventilation will end in two
(2) minutes.
This is an informational message only
(See the LTV
®
1200 or the LTV
®
1150
Ventilator Operator's Manual for an
explanation of this feature).