GB
38
Operation Display
+ In automatic mode input is not possible (see chapter C 3.2.1).
3.3.8 Calculation of the correction factor in automatic mode
In the automatic mode direct calibration is not possible. However, the possibility of filling a preselected quantity
into a measuring vessel and calculating the new correction factor then is given. Following steps are required:
Enter filling volume (see chapter C 3.2.2)
Start filling process (see chapter C 3.2.3)
Change into manual mode at the end of filling process (see chapter C 3.2.1)
Have correction factor displayed from the main memory (see chapter C 3.3.3)
Calculate new correction factor according to the below formula
k
1
: Correction factor from the main memory
k
2
: New correction factor
V
V
: Preselected volume
V
M
: Read volume in measuring vessel
Enter calculated correction factor into the main memory (see chapter C 3.3.2)
k
V
V
k
M
V
2
1
=