6-8
b. Disconnect the valve cartridge from the front of the control module.
c. Optional: Connect the probe’s two lateral vacuum tube set lines to each other and pinch the
pinch valve, if desired, for disposal.
NOTE: Do not remove vacuum tubing from the vacuum canister if vacuum is being used (ON).
d. To remove the probe from the holster, press down on the locking tab on the back of the probe
body. Pull the holster upward about two inches and then backwards to disconnect it from the
probe.
NOTE: If BIOPSY was the last button pressed before removal and the system has not been shut
down or put into standby, the “Check Tissue” screen will be shown. (See Chapter 9:
Service and Troubleshooting for further instruction.)
e. When using the sleeve:
i., ii. Slide the sleeve up the cable, back over the holster.
iii. With two hands, pull the sleeve from the holster.