Setting up and running the Instrument
3. Place the plate in the plate carriage.
4. Verify that the READY light and the correct LED for
your plate type are lit on the keypad.
5. Test that the nozzle-tips height is correct by using
the nozzle-height guide.
6. Press the MODE key twice; the MEMORY LED is
lit. Then select the program you want to run.
7. Press SELECT. LoAd appears in the LED display.
8. Press START. The program dispense volume
appears in the LED display.
9. Press START. The program begins running.
10. Press STOP if you need to stop the program.
Press START if you want to restart from the point
where the program stopped; press RESET to
restart the program from the beginning.
11. When you are nished with the dispensing
operations, return the agent remaining in the
tubes to the supply receptacle or other container.
Then rinse the tubing rst with water and then
with alcohol to dry the tubing, or perform the
appropriate cleaning action (see the maintenance
information in the next section).
12. When you have nished using the instrument,
leave the tubing in a resting position: unfold the
left-hand side of the tubing assembly.